Title of article :
Copper determination in ethanol fuel by differential pulse anodic stripping voltammetry at a solid paraffin-based carbon paste electrode modified with 2-aminothiazole organofunctionalized silica
Author/Authors :
Takeuchi، نويسنده , , Regina M. and Santos، نويسنده , , André L. and Padilha، نويسنده , , Pedro M. and Stradiotto، نويسنده , , Nelson R.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2007
Pages :
7
From page :
771
To page :
777
Abstract :
Solid paraffin-based carbon paste electrodes modified with 2-aminothiazole organofunctionalized silica have been applied to the anodic stripping determination of copper ions in ethanol fuel samples without any sample treatment. The proposed method comprised four steps: (1) copper ions preconcentration at open circuit potential directly in the ethanol fuel sample; (2) exchange of the solution and immediate cathodic reduction of the absorbate at controlled potential; (3) differential pulse anodic stripping voltammetry; (4) electrochemical surface regeneration by applying a positive potential in acid media. Factors affecting the preconcentration, reduction and stripping steps were investigated and the optimum conditions were employed to develop the analytical procedure. Using a preconcentration time of 20 min and reduction time of 120 s at −0.3 V versus Ag/AgClsat a linear range from 7.5 × 10−8 to 2.5 × 10−6 mol L−1 with detection limit of 3.1 × 10−8 mol L−1 was obtained. Interference studies have shown a decrease in the interference effect according to the sequence: Ni > Zn > Cd > Pb > Fe. However, the interference effects of these ions have not forbidden the application of the proposed method. Recovery values between 98.8 and 102.3% were obtained for synthetic samples spiked with known amounts of Cu2+ and interfering metallic ions. The developed electrode was successfully applied to the determination of Cu2+ in commercial ethanol fuel samples. The results were compared to those obtained by flame atomic absorption spectroscopy by using the F-test and t-test. Neither F-value nor t-value have exceeded the critical values at 95% confidence level, confirming that there are no significant differences between the results obtained by both methods.
Keywords :
Ethanol Fuel , Solid paraffin-based carbon paste electrodes , Organofunctionalized silica , copper ions , Differential pulse stripping voltammetry
Journal title :
Talanta
Serial Year :
2007
Journal title :
Talanta
Record number :
1651330
Link To Document :
بازگشت