Title of article
Application of gallium film electrode for elimination of copper interference in anodic stripping voltammetry of zinc
Author/Authors
Katarzyna Tyszczuk، نويسنده , , Katarzyna and Korolczuk، نويسنده , , Mieczyslaw and Grabarczyk، نويسنده , , Malgorzata، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
4
From page
2098
To page
2101
Abstract
For elimination of copper interference in anodic stripping determinations of zinc at mercury and bismuth film electrodes gallium ions are usually added to the supporting electrolyte. In the presented studies novel ex situ formed gallium film electrode was applied for this purpose. The proposed electrode is less toxic than mercury one while the detection limit for zinc was lower than for bismuth film electrode following the same deposition time. The calibration graph for deposition time of 60 s was linear from 5 × 10−8 to 2 × 10−6 mol L−1. The determinations of zinc were carried out from undeaerated solutions. The proposed procedure was applied to zinc determination in certified reference material and tap water sample.
Keywords
Gallium film electrode , Zinc , Determination , Stripping voltammetry
Journal title
Talanta
Serial Year
2007
Journal title
Talanta
Record number
1651948