Author/Authors :
Chen، نويسنده , , Zhao-Peng and Peng، نويسنده , , Zhaofeng and Zhang، نويسنده , , Peng and Jin، نويسنده , , Xue-Fang and Jiang، نويسنده , , Jian-Hui and Zhang، نويسنده , , Xiaobing and Shen، نويسنده , , Guo-Li and Yu، نويسنده , , Ru-Qin، نويسنده ,
Abstract :
A sensitive immunosensor using colloidal gold as electrochemical label is described. In this method, the capture protein was first immobilized on a carbon paste electrode surface through passive adsorption to bind quantitatively with corresponding antigen and colloidal gold labeled antibody to perform a sandwich assay. To detect the amount of the colloidal gold captured on the electrode surface, the colloid was first oxidized electrochemically to produce AuCl4− ions which were adsorbed strongly on the electrode surface. Adsorptive voltammetry was then employed for the determination of the adsorbed AuCl4− ions. A linear relationship between reduction wave peak current and the antigen concentration (human IgG) from 10 to 500 ng/ml is obtained with a detection limit of 4.0 ng/ml.
Keywords :
Immunosensor , Colloidal gold , Adsorptive voltammetry , Electrochemical label