Title of article :
Speciation of iodate and iodide in seawater by non-suppressed ion chromatography with inductively coupled plasma mass spectrometry
Author/Authors :
Chen، نويسنده , , ZuLiang and Megharaj، نويسنده , , Mallavarapu and Naidu، نويسنده , , Ravendra، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2007
Pages :
5
From page :
1842
To page :
1846
Abstract :
A non-suppressed ion chromatography (IC) with inductively coupled plasma mass spectrometry (ICP-MS) has been developed for simultaneous determination of trace iodate and iodide in seawater. An anion-exchange column (G3154A/101, provided by Agilent) was used for the separation of iodate and iodide with an eluent containing 20 mM NH4NO3 at pH 5.6, which reduced the build-up of salts on the sampler and skimmer cones. The influences of competing ion (NO3−) in the eluent on the retention time and detection sensitivity were investigated to give reasonable resolution and detection limits. Linear plots were obtained in a concentration range of 5.0–500 μg/L and the detection limit was 1.5 μg/L for iodate and 2.0 μg/L for iodide. The proposed method was used to determinate iodate and iodide in seawaters without sample pre-treatment with exception of dilution.
Keywords :
Iodate and iodide , Seawater , IC–ICP-MS
Journal title :
Talanta
Serial Year :
2007
Journal title :
Talanta
Record number :
1652782
Link To Document :
بازگشت