Title of article
DRC™ ICP-MS coupled with automated flow injection system with anion exchange minicolumns for determination of selenium compounds in water samples
Author/Authors
Jitmanee، Kritsana نويسنده , , Kritsana and Teshima، نويسنده , , Norio and Sakai، نويسنده , , Tadao and Grudpan، نويسنده , , Kate، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
6
From page
352
To page
357
Abstract
A flow injection system with anion exchange resin minicolumns was coupled with dynamic reaction cell (DRC™) ICP-MS for the determination and speciation of selenite and selenate at sub μg L−1 levels. The charged selenate and uncharged selenite were separated on the first resin column in which only selenate was retained. The unretained selenite was then deprotonated with alkaline solution, and the resulting anionic selenite species was collected on the second column serially connected downstream. By setting a sample loop, total selenium can be determined together with selenite and selenate. The selenium species was eluted by nitric acid and carried to DRC™ ICP-MS for their detection. Using ammonia as reaction gas, the detection of 78Se was improved. The enrichment factor was 20 for 10 mL of sample. The standard deviations (n = 5) of peak heights were 4.9%, 4.1%, and 7.0% for a 5.0 × 10−2 μg L−1 selenite and selenate, and total Se, respectively. The calibration graphs were linear from 2.0 × 10−2 to 1.0 μg L−1 selenite and selenate. And, the linearity for total selenium was good in the range of 10.0 × 10−2 to 1.0 μg L−1. The proposed method has been demonstrated for the application to natural and bottled drinking water samples.
Keywords
DRC™ ICP-MS , Speciation , Inorganic selenium , Anion exchange resin minicolumn , Automated flow injection system , Selenite and selenate
Journal title
Talanta
Serial Year
2007
Journal title
Talanta
Record number
1652943
Link To Document