• Title of article

    Reverse flow injection spectrophotometric determination of iron(III) using chlortetracycline reagent

  • Author/Authors

    Wirat Ruengsitagoon، نويسنده , , Wirat، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    1236
  • To page
    1241
  • Abstract
    A simple reversed flow injection colourimetric procedure for determining iron(III) was proposed. It is based on the reaction between iron(III) with chlortetracycline, resulting in an intense yellow complex with a suitable absorption at 435 nm. A 200 μl chlortetracycline reagent solution was injected into the phosphate buffer stream (flow rate 2.0 ml min−1) which was then merged with iron(III) standard or sample in dilute nitric acid stream (flow rate 1.5 ml min−1). Optimum conditions for determining iron(III) were investigated by univariate method. Under the optimum conditions, a linear calibration graph was obtained over the range 0.5–20.0 μg ml−1. The detection limit (3σ) and the quantification limit (10σ) were 0.10 and 0.82 μg ml−1, respectively. The relatives standard deviation of the proposed method calculated from 12 replicate injections of 2.0 and 10.0 μg ml−1 iron(III) were 0.43 and 0.59%, respectively. The sample throughput was 60 h−1. The proposed method has been satisfactorily applied to the determination of iron(III) in natural waters.
  • Keywords
    Flow Injection , Chlortetracycline , Iron(III) , Reverse flow injection
  • Journal title
    Talanta
  • Serial Year
    2008
  • Journal title
    Talanta
  • Record number

    1654064