Title of article :
Simultaneous determination of bromide and nitrate in contaminated waters by ion chromatography using amperometry and absorbance detectors
Author/Authors :
Tirumalesh، نويسنده , , K.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2008
Pages :
7
From page :
1428
To page :
1434
Abstract :
This study describes a new ion chromatography method using a low-capacity anion exchange column with amperometric and absorbance detection for rapid and simultaneous determination of Br− and NO3− in contaminated waters where one of these ions is present in excess compared to other. The use of two detectors overcomes the problem of baseline separation for Br− and NO3− for accurate quantification, which was commonly encountered when using a low-capacity anion exchange column and suppressed conductivity detection mode. The method achieved accurate quantification of these two ions without requirement of baseline separation. The accuracy of 2.8% for NO3− was determined using a quality control sample obtained from UN GEMS/Water PE Study No. 6. The detection limits for Br− and NO3− were 20 and 6 μg l−1 (25 μl sample), respectively. Linearity of these two ions was over three orders of magnitude with a correlation coefficient >0.998. The influence of potential interfering ions was also studied followed by the determination of Br− and NO3− in seawater, unsaturated zone water, soil extract and groundwater.
Keywords :
Ion chromatography , Bromide , nitrate , Amperometry , UV absorbance detector
Journal title :
Talanta
Serial Year :
2008
Journal title :
Talanta
Record number :
1654166
Link To Document :
بازگشت