• Title of article

    Spectrophotometric determination of chloride in waters using a multisyringe flow injection system

  • Author/Authors

    Maya، نويسنده , , Fernando and Estela، نويسنده , , José Manuel and Cerdà، نويسنده , , Vيctor، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    1534
  • To page
    1538
  • Abstract
    A multisyringe flow injection system (MSFIA) with spectrophotometric detection is proposed as a fast, robust and low-reagent consumption system for the determination of chloride (Cl−) in waters. The system is based in the classic reaction of Cl− with Fe3+ and Hg(SCN)2, but due to the hazardous properties of this last reagent, the proposed methodology has been developed with the aim to minimize the consumption of this one, consuming less than 0.05 mg of Hg for a Cl− determination, being the system of this type with the lowest Hg consumption. The linear working range was between 1 and 40 mg L−1 Cl− and the detection limit was 0.2 mg L−1 Cl−. The repeatability (RSD) was 0.8% for a 10 mg L−1 Cl− solution, and the injection throughput was 130 h−1. The proposed system is compared with other chloride monitoring flow systems, this comparison is realized with a point of view of the equilibrium between the obtained analytical features and produced residues toxicity. The proposed system was applied to the determination of Cl− in mineral, tap and well water.
  • Keywords
    Multisyringe flow injection analysis (MSFIA) , chloride , Mercury-thiocyanate , Water samples
  • Journal title
    Talanta
  • Serial Year
    2008
  • Journal title
    Talanta
  • Record number

    1654215