Title of article
Determination of chloride, chlorate and perchlorate by PDMS microchip electrophoresis with indirect amperometric detection
Author/Authors
Li، نويسنده , , Xin-Ai and Zhou، نويسنده , , Dongmei and Xu، نويسنده , , Jing-Juan and Chen، نويسنده , , Hong-Yuan، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2008
Pages
6
From page
157
To page
162
Abstract
In this work, chloride, chlorate and perchlorate are fast separated on PDMS microchip and detected via in-channel indirect amperometric detection mode. With PDMS/PDMS microchip treated by oxygen plasma, anions chloride (Cl−), chlorate (ClO3−), and perchlorate (ClO4−) are separated within 35 s. Some parameters including buffer salt concentration, buffer pH, separation voltage and detection potential are investigated in detail. The separation conditions using 15 mM (pH 6.12) of 2-(N-morpholino)ethanesulfonic acid (MES) + l-histidine (l-His) as running buffer, −2000 V as separation voltage and 0.7 V as detection potential are optimized. Under this condition, the detection limits of Cl−, ClO3−, and ClO4− are 1.9, 3.6, and 2.8 μM, respectively.
Keywords
Poly(dimethylsiloxane) , Chlorine-containing anions , Microchip capillary electrophoresis , Oxygen plasma treatment , Carbon fiber electrode , Indirect amperometric detection