Title of article :
Validation of a fluorimetric assay for 4-aminophenol in paracetamol formulations
Author/Authors :
Dejaegher، نويسنده , , B. P. Bloomfield، نويسنده , , M.S. and Smeyers-Verbeke، نويسنده , , J. and Vander Heyden، نويسنده , , Y.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2008
Abstract :
4-Aminophenol (4-AP) is the primary degradation product of paracetamol (PARA). According to the European Pharmacopoeia, 50 ppm 4-AP/PARA is the specification limit of 4-aminophenol in paracetamol drug substance. For drug products, often higher specification limits, such as 1000 ppm 4-AP/PARA are applied. This paper describes a fluorimetric method to quantify the low amount of this degradant (50 ppm) in a pharmaceutical preparation, i.e. in paracetamol tablets. The fluorimetric method was validated and the linearity, precision, trueness, range, limit of detection and limit of quantification were determined. They were found acceptable to assay the low amounts of 4-aminophenol in paracetamol tablets.
Keywords :
4-Aminophenol , Fluorimetry , Method validation