• Title of article

    Electrochemical behavior of colchicine using graphite-based screen-printed electrodes

  • Author/Authors

    Bodoki، نويسنده , , E. and Laschi، نويسنده , , Serena and Palchetti، نويسنده , , Ilaria and S?ndulescu، نويسنده , , R. and Mascini، نويسنده , , M.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2008
  • Pages
    7
  • From page
    288
  • To page
    294
  • Abstract
    Miniaturized electrochemical graphite based sensor for the analysis of colchicine, along with its detailed construction was described. ectrochemical behavior of colchicine, both by cyclic and differential pulse voltammetry was investigated, showing an irreversible reduction and oxidation mechanism. The effect of several different electrochemical mediators was studied in order to decrease the oxidation potential of colchicine, but none of them showed a favorable electron transfer between the alkaloid and the electrodeʹs surface. The influences of different biomolecules (calf thymus DNA, bovine serum albumin and β-tubulin) over colchicineʹs electrochemical behavior were also presented. antitative determinations the anodic differential pulse voltammetric method in H3PO4/HClO4 0.01 M (pH 2.05) was selected, evaluating one of the oxidation peaks of colchicine at +970 mV versus Ag pseudo-reference. The electrochemical scanning parameters were optimized by an experimental design using response surface modelling. The method was validated according to ICH guidelines in the concentration range 85–1200 ng mL−1 (R2 = 0.9966, n = 7) colchicine in terms of linearity, accuracy, limits of detection (LOD) (41 ng mL−1), limits of quantification (LOQ) and fidelity and it was successfully applied to the determination of colchicine in tablets, without the interference of the excipients. udied method showed the same sensitivity as the more complex HPLC procedure and micellar electrokinetic chromatography with spectrophotometric detection.
  • Keywords
    Screen-printed electrodes , Differential pulse voltammetry , Method validation , Colchicine
  • Journal title
    Talanta
  • Serial Year
    2008
  • Journal title
    Talanta
  • Record number

    1655179