Title of article
Use of a capacitance measurement device for surrogate noncontact conductance measurement
Author/Authors
Takeuchi، نويسنده , , Masaki and Li، نويسنده , , Qingyang and Yang، نويسنده , , Bingcheng and Dasgupta، نويسنده , , Purnendu K. and Wilde، نويسنده , , Vincent E.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2008
Pages
4
From page
617
To page
620
Abstract
A capacitance to digital converter (AD7746) is used in the same mode as noncontact conductance detectors. The detector output is linearly proportional to specific conductance (σ) at low σ values but becomes nonlinear and reaches a plateau value at σ ∼>0.75 mS/cm, regardless of the nature of the electrolyte. For all applications at sub- to low-mM concentrations, the device, available as an evaluation board, provides a very affordable nothing-else-required means of contactless surrogate conductivity detection from capillary scale to larger bore conduits. For the same measurement cell volume, the detector provides virtually the same limits of detection (LODs) as a standard galvanic contact conductivity detector in conventional scale suppressed conductometric ion chromatography. The detection limits deteriorate as the conduit inner diameter deceases.
Keywords
Contactless conductivity , capacitance , Ion chromatography
Journal title
Talanta
Serial Year
2008
Journal title
Talanta
Record number
1655350
Link To Document