Title of article :
Direct determination of ammoniacal nitrogen in water samples using corona discharge ion mobility spectrometry
Author/Authors :
Jafari، نويسنده , , M.T. and Khayamian، نويسنده , , T.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2008
Pages :
5
From page :
1189
To page :
1193
Abstract :
In this study, direct determination of ammoniacal nitrogen residues in water samples using corona discharge ion mobility spectrometry (CD-IMS) was investigated. Pyridine was used as an alternate reagent gas to enhance selectivity and sensitivity of the method. The results indicate that the limit of detection (LOD) was about 9.2 × 10−3 μg mL−1 and the linear dynamic range was obtained from 0.03 to 2.00 μg mL−1. The relative standard deviation was about 11%. Furthermore, this method was successfully applied to the direct determination of ammoniacal nitrogen in river and tap water samples and the results were compared with the Nessler method. The comparison of the results validates the potential of the proposed method as an alternative technique for the analysis of the ammoniacal nitrogen in water samples.
Keywords :
Corona discharge , Ammoniacal nitrogen , water , Ion mobility spectrometry
Journal title :
Talanta
Serial Year :
2008
Journal title :
Talanta
Record number :
1655664
Link To Document :
بازگشت