Title of article :
A LC–MS method allowing the analysis of HMX and RDX present at the picogram level in natural aqueous samples without a concentration step
Author/Authors :
Vigneau، نويسنده , , Olivier and Machuron-Mandard، نويسنده , , Xavier، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Abstract :
The introduction of chloroform into the nebulising gas of a LC/MS electrospray interface (ESI), in a perfectly controlled way, leads to the formation of intense adducts ([M+Cl]−) when a mobile phase containing HMX (1,3,5,7-tetranitro-1,3,5,7-tetrazacyclooctane or octogen) and RDX (1,3,5-trintro-1,3,5-triazacyclohexane or hexogen) is eluted. This LC/MS method allows the direct analysis of aqueous samples containing HMX and RDX at the pictogram level without a concentration step. The method is used to determine HMX and RDX concentrations in ground water samples from a military site.
Keywords :
ESI , LC/MS , Explosive , HMX , RDX