Title of article :
Comparison of tungsten coil electrothermal vaporization and thermospray sample introduction methods for flame furnace atomic absorption spectrometry
Author/Authors :
Wu، نويسنده , , Peng and Zhang، نويسنده , , Yunchang and Liu، نويسنده , , Rui and Lv، نويسنده , , Yi and Hou، نويسنده , , Xiandeng، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
5
From page :
1778
To page :
1782
Abstract :
Flame furnace atomic absorption spectrometry (FF-AAS) is a newly developed flame atomic absorption spectrometric technique based on arranging a flame furnace onto the top of the flame burner head. In this fundamental investigation, 25 elements were carefully tested by using either thermospray FF-AAS or tungsten coil electrothermal vaporization FF-AAS, of which 15 volatile and semi-volatile elements (Cd, Tl, Ag, Pb, Zn, Hg, Cu, Sb, Bi, Te, In, As, Se, Sn and Au) exhibited better limits of detection compared to those by conventional FAAS; however, non-volatile or refractory elements (Fe, Co, Ni, Cr, Mn, Pd, Pt, Al, Be and V) showed inferior sensitivities by the proposed methods.
Keywords :
Electrothermal vaporization , Flame furnace , Atomic absorption spectrometry , Thermospray , Tungsten coil