Title of article :
A single standard calibration module for flow analysis systems based on solenoid microdevices
Author/Authors :
Tymecki، نويسنده , , ?ukasz and Strzelak، نويسنده , , Kamil and Koncki، نويسنده , , Robert، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
6
From page :
205
To page :
210
Abstract :
Only two computer-controlled microsolenoid devices, namely two micropumps or one micropump and one microvalve, are sufficient for the construction of on-line dilution modules useful in several flow analytical systems for the calibration using single standard. Three simple constructions of such modules were tested and compared. The most promising is the one based on the concept of a microvalve controlling dilution ratio of the standard and a solenoid micropump playing a double role: solution pumping device and mixing segments homogenizer. All investigated modules were tested with paired emitter detector diode (PEDD) as photometric flow-through detector and bromothymol blue as a model analyte. The best module was implemented into more advanced flow-injection system dedicated for optical detection of alkaline phosphatase activity using UV-PEDD-based flow-through detector for the enzyme reaction product.
Keywords :
Solenoid device , PEDD , Calibration , Flow Analysis
Journal title :
Talanta
Serial Year :
2009
Journal title :
Talanta
Record number :
1658086
Link To Document :
بازگشت