Title of article
Development of an automatic solid phase extraction and liquid chromatography mass spectrometry method by using a monolithic column for the analysis of Cyclosporin A in human plasma
Author/Authors
Alvarez، نويسنده , , Covadonga and Wainer، نويسنده , , Irving W.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2009
Pages
4
From page
280
To page
283
Abstract
A sensitive and specific and automated liquid chromatography–electrospray mass spectrometric (LC–ESI-MS) assay for the quantification of Cyclosporin A in human plasma was developed. Following a simple protein precipitation step, the supernatant was extracted on-line and directly injected into the system LC–ESI-MS. A relatively new type of monolithic column consisting of a silica rod with bimodal pore structure was used to achieve a retention time of 2.4 min with a very low backpressure at a flow rate of 1 ml/min. The assay was linear from 0.050 to 1.000 μg/ml. The mean recovery was 91%. The mean inter-day and intra-day precisions were 1.85% and 2.83%, respectively. The combination of the automated solid phase extraction and the low retention time achieved with this columns increase the throughput and decrease the time of analysis of each sample. This technology is useful in order to improve the efficiency of the bioanalytical studies.
Keywords
cyclosporin A , Monolithic columns , LC–ESI-MS
Journal title
Talanta
Serial Year
2009
Journal title
Talanta
Record number
1658118
Link To Document