• Title of article

    Simple and rapid voltammetric determination of morphine at electrochemically pretreated glassy carbon electrodes

  • Author/Authors

    Li، نويسنده , , Fei and Song، نويسنده , , Jixia and Gao، نويسنده , , Dongmei and Zhang، نويسنده , , Qixian and Han، نويسنده , , Dongxue and Niu، نويسنده , , Li، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    845
  • To page
    850
  • Abstract
    A simple and rapid method for morphine detection has been described based on electrochemical pretreatment of glassy carbon electrode (GCE) which was treated by anodic oxidation at 1.75 V, following potential cycling in the potential range from 0 V to 1.0 V vs. Ag|AgCl reference electrode. The sensitivity for morphine detection was improved greatly and the detection limit was 0.2 μM. The reproducibility of the voltammetric measurements was usually less than 3% RSD for six replicate measurements. Moreover, this method could readily discriminate morphine from codeine. And an electrochemical detection of morphine in spiked urine sample was succeeded with satisfactory results.
  • Keywords
    Morphine , Electrochemical pretreatment , Glassy carbon electrode , Voltammetric detection
  • Journal title
    Talanta
  • Serial Year
    2009
  • Journal title
    Talanta
  • Record number