• Title of article

    Integrated multi-sensor chip with photocured polymer membranes containing copolymerised plasticizer for direct pH, potassium, sodium and chloride ions determination in blood serum

  • Author/Authors

    Abramova، نويسنده , , Natalia and Ipatov، نويسنده , , Andrey and Levichev، نويسنده , , Sergey and Bratov، نويسنده , , Andrey، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    984
  • To page
    989
  • Abstract
    An analytical system based on a sensor array with ion-selective field effect transistors (ISFETs) monolithically integrated in one chip covered with photocured polymer membranes containing copolymerised plasticizer and a sequential injection analysis (SIA) is shown to offer an automation of the analysis of blood serum components. For sequential injection system a custom made dual channel flow cell for the sensor array was developed. Optimisation of ion-sensitive membrane characteristics and calibration solution compositions were carried out. The system was used to analyze sodium, potassium, chloride ion contents in blood serum samples. The precision of the ion determination in samples was typical for potentiometic method with standard deviation of about 3–5%.
  • Keywords
    ISFET sensor array , Photocured polymer membranes , Sequential injection analysis , Blood analysis
  • Journal title
    Talanta
  • Serial Year
    2009
  • Journal title
    Talanta
  • Record number