Title of article
On-line solid phase extraction system using 1,10-phenanthroline immobilized on surfactant coated alumina for the flame atomic absorption spectrometric determination of copper and cadmium
Author/Authors
Shabani، نويسنده , , Ali Mohammed Haji and Dadfarnia، نويسنده , , Shayessteh and Dehghani، نويسنده , , Zahra، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2009
Pages
5
From page
1066
To page
1070
Abstract
An on-line solid phase extraction (SPE) preconcentration system coupled to flame atomic absorption spectrometer (FAAS) was developed for determination of copper and cadmium at μg L−1 level. The method is based on the on-line retention of copper and cadmium on a microcolumn of alumina modified with sodium dodecyl sulfate (SDS) and 1,10-phenanthroline and subsequent elution with ethanol and determination by FAAS. The effect of chemical and flow variables that could affect the performance of the system was investigated. The relative standard deviation (n = 6) at 20 μg L−1 level for copper and cadmium were 1.4 and 2.2% and the corresponding limits of detection (based on 3σ) were 0.04 and 0.14 μg L−1, respectively. The method was successfully applied to determination of copper and cadmium in human hair and water samples.
Keywords
Cadmium , Copper , Flow Injection , 1 , Atomic absorption spectrometry , 10-Phenanthroline
Journal title
Talanta
Serial Year
2009
Journal title
Talanta
Record number
1658503
Link To Document