Title of article :
Uncertainty of high resolution inductively coupled plasma mass spectrometry based aerosol measurements
Author/Authors :
S.G and Yenisoy-Karakas، نويسنده , , Serpil and Gaga، نويسنده , , Eftade O. and Cankur، نويسنده , , Oktay and Karaka?، نويسنده , , Duran، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
8
From page :
1298
To page :
1305
Abstract :
Simultaneous and multi-elemental method was optimized to determine 35 elements in 274 coarse and fine aerosol samples by using HR-ICP-MS. The procedures were validated. Measurement uncertainties of all elements including sampling uncertainties were calculated by applying bottom-up approach. The average recoveries obtained for each element ranged between 79% and 129% using NIST SRM 1648, urban dust. The calculated uncertainties of the analytical methods were between 2.9% and 18% for both sample types. The major contributions to the uncertainty budget come from the calibration curves, repeatability and volume of air.
Keywords :
Elements , aerosols , Method validation , Microwave digestion , uncertainty , High resolution inductively coupled plasma mass spectrometry
Journal title :
Talanta
Serial Year :
2009
Journal title :
Talanta
Record number :
1658605
Link To Document :
بازگشت