Title of article
Six orders of magnitude dynamic range in capillary electrophoresis with ultrasensitive laser-induced fluorescence detection
Author/Authors
Whitmore، نويسنده , , Colin D. and Essaka، نويسنده , , David and Dovichi، نويسنده , , Norman J.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2009
Pages
5
From page
744
To page
748
Abstract
An ultrasensitive laser-induced fluorescence detector was used with capillary electrophoresis for the study of 5-carboxy-tetramethylrhodamine. The raw signal from the detector provided roughly three orders of magnitude dynamic range. The signal saturated at high analyte concentrations due to the dead time associated with the single-photon counting avalanche photodiode employed in the detector. The signal can be corrected for the detector dead time, providing an additional order of magnitude dynamic range. To further increase dynamic range, two fiber-optic beam-splitters were cascaded to generate a primary signal and two attenuated signals, each monitored by a single-photon counting avalanche photodiode. The combined signals from the three photodiodes are reasonably linear from the concentration detection limit of 3 pM to 10 μM, the maximum concentration investigated, a range of 3,000,000. Mass detection limits were 150 yoctomoles injected onto the capillary.
Keywords
Capillary electrophoresis , dynamic range , Laser induced fluorescence , Linear range
Journal title
Talanta
Serial Year
2009
Journal title
Talanta
Record number
1659036
Link To Document