• Title of article

    Six orders of magnitude dynamic range in capillary electrophoresis with ultrasensitive laser-induced fluorescence detection

  • Author/Authors

    Whitmore، نويسنده , , Colin D. and Essaka، نويسنده , , David and Dovichi، نويسنده , , Norman J.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2009
  • Pages
    5
  • From page
    744
  • To page
    748
  • Abstract
    An ultrasensitive laser-induced fluorescence detector was used with capillary electrophoresis for the study of 5-carboxy-tetramethylrhodamine. The raw signal from the detector provided roughly three orders of magnitude dynamic range. The signal saturated at high analyte concentrations due to the dead time associated with the single-photon counting avalanche photodiode employed in the detector. The signal can be corrected for the detector dead time, providing an additional order of magnitude dynamic range. To further increase dynamic range, two fiber-optic beam-splitters were cascaded to generate a primary signal and two attenuated signals, each monitored by a single-photon counting avalanche photodiode. The combined signals from the three photodiodes are reasonably linear from the concentration detection limit of 3 pM to 10 μM, the maximum concentration investigated, a range of 3,000,000. Mass detection limits were 150 yoctomoles injected onto the capillary.
  • Keywords
    Capillary electrophoresis , dynamic range , Laser induced fluorescence , Linear range
  • Journal title
    Talanta
  • Serial Year
    2009
  • Journal title
    Talanta
  • Record number