• Title of article

    Development and validation by accuracy profile of a method for the analysis of monoterpenes in indoor air by active sampling and thermal desorption-gas chromatography-mass spectrometry

  • Author/Authors

    Marlet، نويسنده , , Christelle and Lognay، نويسنده , , Georges، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2010
  • Pages
    10
  • From page
    1230
  • To page
    1239
  • Abstract
    The technique of thermal desorption (TD)-GC/MS was evaluated for the measurement of monoterpenes in indoor air. The validation strategy was intentionally oriented towards routine use and the reliability of the method rather than extreme performance. For this reason, validation by accuracy profile was chosen. The accuracy profile procedure, which is based on the concept of total error (bias + standard deviation), guarantees that a known proportion of future results obtained with the method will be within acceptance limits. For all the compounds tested in the present study, α-pinene, α-terpineol, β-pinene, d-limonene, Δ3-carene, camphene, 1,8-cineole, p-cymene, linalool, but not in the case of carvone, the accuracy profile procedure established that at least 95% of the future results obtained would be within the ±15% acceptance limits of the validated method over the whole defined concentration range. Other parameters, such as selectivity, recovery, repeatability, stability of the molecules of interest and the effect of temperature, were also determined. The performance of the described method was finally evaluated by the analysis of indoor air from new timber frame constructions.
  • Keywords
    monoterpenes , thermal desorption , Accuracy profile , Tenax TA , Air sampling , wood construction
  • Journal title
    Talanta
  • Serial Year
    2010
  • Journal title
    Talanta
  • Record number

    1660588