Title of article :
Development and validation by accuracy profile of a method for the analysis of monoterpenes in indoor air by active sampling and thermal desorption-gas chromatography-mass spectrometry
Author/Authors :
Marlet، نويسنده , , Christelle and Lognay، نويسنده , , Georges، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2010
Pages :
10
From page :
1230
To page :
1239
Abstract :
The technique of thermal desorption (TD)-GC/MS was evaluated for the measurement of monoterpenes in indoor air. The validation strategy was intentionally oriented towards routine use and the reliability of the method rather than extreme performance. For this reason, validation by accuracy profile was chosen. The accuracy profile procedure, which is based on the concept of total error (bias + standard deviation), guarantees that a known proportion of future results obtained with the method will be within acceptance limits. For all the compounds tested in the present study, α-pinene, α-terpineol, β-pinene, d-limonene, Δ3-carene, camphene, 1,8-cineole, p-cymene, linalool, but not in the case of carvone, the accuracy profile procedure established that at least 95% of the future results obtained would be within the ±15% acceptance limits of the validated method over the whole defined concentration range. Other parameters, such as selectivity, recovery, repeatability, stability of the molecules of interest and the effect of temperature, were also determined. The performance of the described method was finally evaluated by the analysis of indoor air from new timber frame constructions.
Keywords :
monoterpenes , thermal desorption , Accuracy profile , Tenax TA , Air sampling , wood construction
Journal title :
Talanta
Serial Year :
2010
Journal title :
Talanta
Record number :
1660588
Link To Document :
بازگشت