Title of article :
Industrial application of green chromatography—I. Separation and analysis of niacinamide in skincare creams using pure water as the mobile phase
Author/Authors :
Yang، نويسنده , , Yu and Strickland، نويسنده , , Zackary and Kapalavavi، نويسنده , , Brahmam and Marple، نويسنده , , Ronita and Gamsky، نويسنده , , Chris، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2011
Abstract :
In this work, chromatographic separation of niacin and niacinamide using pure water as the sole component in the mobile phase has been investigated. The separation and analysis of niacinamide have been optimized using three columns at different temperatures and various flow rates. Our results clearly demonstrate that separation and analysis of niacinamide from skincare products can be achieved using pure water as the eluent at 60 °C on a Waters XTerra MS C18 column, a Waters XBridge C18 column, or at 80 °C on a Hamilton PRP-1 column. The separation efficiency, quantification quality, and analysis time of this new method are at least comparable with those of the traditional HPLC methods. Compared with traditional HPLC, the major advantage of this newly developed green chromatography technique is the elimination of organic solvents required in the HPLC mobile phase. In addition, the pure water chromatography separations described in this work can be directly applied in industrial plant settings without further modification of the existing HPLC equipment.
Keywords :
High-temperature water chromatography , HPLC , Niacin , Niacinamide , 4-Acetamidophenol , Skincare cream , Industrial application , Procter & , Gamble , Pure water chromatography , Subcritical water chromatography