Title of article
Separation and preconcentration of Pb(II) using ionic liquid-modified silica and its determination by flame atomic absorption spectrometry
Author/Authors
Ayata، نويسنده , , Sevda and Bozkurt، نويسنده , , Serap Seyhan and Ocakoglu، نويسنده , , Kasim، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2011
Pages
4
From page
212
To page
215
Abstract
A new method based on microcolumn packed with ionic liquid-modified silica combined with flame atomic absorption spectrometry has been developed for the determination of lead in environmental samples. Several factors influencing the preconcentration efficiency of lead and its subsequent determination, such as pH of the sample, flow rate, mass of ionic liquid, and interfering effect, have been investigated. Lead could be quantitatively retained by ionic liquid-modified silica in the pH range of 5–7, and then eluted completely with 3.0 mL 1.0 mol L−1 HCl. The detection limit of this method for lead was 0.7 μg L−1 with preconcentration factor of 185, and the relative standard deviation (RSD) was 4.2% at 0.1 μg mL−1 Pb(II). This method has been applied for the determination of trace amount of lead in NIST standard reference material 2709 (San Joaquin Soil) and river water samples with satisfactory results.
Keywords
silica , Flame AAS , Ionic liquid , Lead , Solid phase extraction
Journal title
Talanta
Serial Year
2011
Journal title
Talanta
Record number
1661914
Link To Document