• Title of article

    Flow injection analysis combined with a hydrothermal flow reactor: Application to kinetic determination of trace amounts of iridium using a water-soluble porphyrin

  • Author/Authors

    Kawamura، نويسنده , , Kunio and Ikoma، نويسنده , , Keisuke and Igarashi، نويسنده , , Shukuro and Hisamoto، نويسنده , , Hideaki and Yao، نويسنده , , Toshio، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    1318
  • To page
    1322
  • Abstract
    A new type of flow injection analysis (FIA) system combined with an extremely high temperature reactor, namely hydrothermal flow injection analysis (HT-FIA), has been successfully constructed for the first time. Fundamental characteristics of HT-FIA system, such as limit temperature, pressure, and flow rate, were examined as an analytical tool. To demonstrate the potential of HT-FIA, the catalytic activity of Ir(IV) for the degradation of a water-soluble porphyrin, 5,10,15,20-tetraphenyl-21H,23H-porphinetetrasulfonic acid (TPPS), was applied for the determination of trace amounts of Ir(IV). Although the indicator reaction is very slow at room temperature, HT-FIA system enables to accelerate the reaction. A linear calibration curve was acquired at 10−8 M level of Ir(IV) and the interferences of platinum group metal ions were examined. The detection limit of Ir(IV) was 5.8 × 10−9 M and a fairly high-throughput analysis, of which more than 30 samples can be analyzed within 80 min, was achieved.
  • Keywords
    iridium , Water-soluble porphyrin , Kinetic determination , Hydrothermal reaction , UV–VIS spectrophotometry
  • Journal title
    Talanta
  • Serial Year
    2011
  • Journal title
    Talanta
  • Record number