Title of article :
Simultaneous determination of N-acetyl-p-aminophenol and p-aminophenol with poly(3,4-ethylenedioxythiophene) modified glassy carbon electrode
Author/Authors :
Mehretie، Solomon نويسنده Department of Chemistry, Addis Ababa University, P.O. Box 1176, Addis Ababa, Ethiopia , , Solomon and Admassie، نويسنده , , Shimelis and Hunde، نويسنده , , Tadele and Tessema، نويسنده , , Merid and Solomon، نويسنده , , Theodros، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2011
Abstract :
A sensitive and selective method was developed for the determination of N-acetyl-p-aminophenol (APAP) and p-aminophenol (PAP) using poly(3,4-ethylenedioxythiophene) (PEDOT)-modified glassy carbon electrode (GCE). Cyclic voltammetry and differential pulse voltammetry were used to investigate the electrochemical reaction of APAP and PAP at the modified electrode. Both APAP and PAP showed quasireversible redox reactions with formal potentials of 367 mV and 101 mV (vs. Ag/AgCl), respectively, in phosphate buffer solution of pH 7.0. The significant peak potential difference (266 mV) between APAP and PAP enabled the simultaneous determination both species based on differential pulse voltammetry. The voltammetric responses gave linear ranges of 1.0 × 10−6–1.0 × 10−4 mol L−1 and 4.0 × 10−6–3.2 × 10−4 mol L−1, with detection limits of 4.0 × 10−7 mol L−1 and 1.2 × 10−6 mol L−1 for APAP and PAP, respectively. The method was successfully applied for the determination of APAP and PAP in pharmaceutical formulations and biological samples.
Keywords :
N-acetyl-p-aminophenol (APAP) , 4-ethylenedioxythiophene) , Poly(3 , p-Aminophenol (PAP) , glassy carbon , Voltammetry