Title of article
A practical method for the determination of sulphur in coal samples by high-resolution continuum source flame atomic absorption spectrometry
Author/Authors
Baysal، نويسنده , , A. and Akman، نويسنده , , S.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2011
Pages
4
From page
2662
To page
2665
Abstract
Sulphur in coal was determined using a high-resolution continuum source flame atomic absorption spectrophotometer (HR-CS-FAAS) with actylene/air flame. The C–S absorption band at 258.056 nm was found the most suitable analytical line with respect to sensitivity and spectral interferences. The instrumental parameters were optimized. The coal samples were dried and dissolved using microwave-assisted digestion technique. The validity of the method was tested using standard reference material and certified values were found in the limits of 95% confidence level. Since the concentrations of matrix elements of coal other than carbon are low enough not to cause any spectral interferences, the linear calibration method was applied in all quantifications without any problem. The calibration standards were prepared in sulphuric acid. The method was accurate, fast, simple and sensitive. The limit of detection (LOD, 3δ, N = 10) and the limit of quantification (LOQ, 10δ, N = 10) were found to be 0.01 and 0.03% (w/w), respectively. The sulphur concentrations of various kinds of the coal samples received around Turkey were determined. The sulphur contents of the coal samples were ranged from ≤LOQ to 1.2%.
Keywords
Sulphur , High resolution continuum source atomic absorption spectrometry , Molecular absorption , Coal
Journal title
Talanta
Serial Year
2011
Journal title
Talanta
Record number
1663523
Link To Document