Title of article :
Uncertainty of ozone measurements with the primary standard reference photometer (SRP45)
Author/Authors :
Leiva G.، نويسنده , , Manuel A. and Araya C.، نويسنده , , Consuelo and Mancilla، نويسنده , , Carlos and Seguel، نويسنده , , Rodrigo and Norris، نويسنده , , James E.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2011
Abstract :
A comparison of the ozone primary reference standard photometer serial number 45 (SRP45) against the National Institute of Standards and Technology (NIST) instruments, serial number 0 (SRP0) and 2 (SRP2), has been performed in order to establish the traceability and comparability of ozone measurements made by the Chilean atmospheric science community. A complete uncertainty budget was developed for SRP45, using a GUM approach. The results of the comparisons allow us to conclude that SRP45, SRP0 and SRP2 are comparable according to internationals criteria over an ozone mole fraction range of 0 nmol mol−1 to at least 500 nmol mol−1. The official result for the validation of SRP45 is x o z o n e S R P 45 = [ 0.013 + 0.99806 x o z o n e S R P 2 ] nmol mol − 1 with an expanded uncertainty of 2 × ( 0.27 ) 2 + ( 1.18 × 10 − 2 × x o z o n e ) 2 from 0 to 500 nmol mol−1.
Keywords :
SRP , Primary standard reference photometer , Air quality , Uncertainty evaluation , GUM