Influence of key parameters on the lower detection limit and response function of solvent polymeric membrane ion-selective electrodes
Author/Authors
Ceresa، نويسنده , , Alan and Sokalski، نويسنده , , Tomasz and Pretsch، نويسنده , , Ernِ، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
7
From page
70
To page
76
Abstract
The lower detection limit of conventional polymeric membrane ion-selective electrodes (ISEs) is given by the release of measuring ions from the sensor membrane into the aqueous sample. It can be greatly improved by using inner solutions of appropriate composition. Recently, steady-state model calculations have been used to predict the influence of various parameters on the response function and the lower detection limit. Among these, effects of the composition of the inner solution, the presence of interfering ions in the sample, the use of a flow-through cell, the polymer content and thickness of the ISE membrane, and the concentration of the membrane components have been investigated in the present work. All observed trends agree fully with the theoretically predicted ones.
Keywords
Polymer membrane electrodes , Membrane composition , Potentiometric sensors , response function , Lower detection limit , Flow-through cell