Title of article :
Development of an automated on-line electrochemical chlorite ion sensor
Author/Authors :
Myers، نويسنده , , John N. and Steinecker، نويسنده , , William H. and Sandlin، نويسنده , , Zechariah D. and Cox، نويسنده , , James A. and Gordon، نويسنده , , Gilbert E. Pacey، نويسنده , , Gilbert E.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2012
Pages :
5
From page :
227
To page :
231
Abstract :
A sensor system for the automatic, in-line, determination of chlorite ion is reported. Electroanalytical measurements were performed in electrolyte-free liquids by using an electrochemical probe (EC), which enables in-line detection in high-resistance media such as disinfected water. Cyclic voltammetry scan rate studies suggest that the current arising from the oxidation of chlorite ion at an EC probe is mass-transfer limited. By coupling FIA with an EC probe amperometric cell, automated analysis was achieved. ensor is intended to fulfill the daily monitoring requirements of the EPA DBP regulations for chlorite ion. Detection limits of 0.02–0.13 mg/L were attained, which is about one order of magnitude below the MRDL. The sensor showed no faradaic signal for perchlorate, chlorate, or nitrate. The lifetime and stability of the sensor were investigated by measuring calibration curves over time under constant-flow conditions. Detection limits of <0.1 mg/L were repeatedly achieved over a period of three weeks.
Keywords :
Electrochemical probe , Chlorite ion , Flow injection analysis
Journal title :
Talanta
Serial Year :
2012
Journal title :
Talanta
Record number :
1665474
Link To Document :
بازگشت