Title of article :
Automated total and radioactive strontium separation and preconcentration in samples of environmental interest exploiting a lab-on-valve system
Author/Authors :
Rodrيguez، نويسنده , , Rogelio and Avivar، نويسنده , , Jessica and Ferrer، نويسنده , , Laura G. Leal، نويسنده , , Luz O. and Cerdà، نويسنده , , Victor، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2012
Abstract :
A novel lab-on-valve system has been developed for strontium determination in environmental samples. Miniaturized lab-on-valve system potentially offers facilities to allow any kind of chemical and physical processes, including fluidic and microcarrier bead control, homogenous reaction and liquid–solid interaction. A rapid, inexpensive and fully automated method for the separation and preconcentration of total and radioactive strontium, using a solid phase extraction material (Sr-Resin), has been developed. Total strontium concentrations are determined by ICP-OES and 90Sr activities by a low background proportional counter. The method has been successfully applied to different water samples of environmental interest. The proposed system offers minimization of sample handling, drastic reduction of reagent volume, improvement of the reproducibility and sample throughput and attains a significant decrease of both time and cost per analysis. The LLD of the total Sr reached is 1.8 ng and the minimum detectable activity for 90Sr is 0.008 Bq. The repeatability of the separation procedure is 1.2% (n = 10).
Keywords :
strontium , LOV , Environmental samples , Solid phase extraction , ICP-OES , Low background proportional counter