Combined corona discharge and UV photoionization source for ion mobility spectrometry
Author/Authors
Bahrami، نويسنده , , Hamed and Tabrizchi، نويسنده , , Mahmoud، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2012
Pages
6
From page
400
To page
405
Abstract
An ion mobility spectrometer is described which is equipped with two non-radioactive ion sources, namely an atmospheric pressure photoionization and a corona discharge ionization source. The two sources cannot only run individually but are additionally capable of operating simultaneously. For photoionization, a UV lamp was mounted parallel to the axis of the ion mobility cell. The corona discharge electrode was mounted perpendicular to the UV radiation. The total ion current from the photoionization source was verified as a function of lamp current, sample flow rate, and drift field. Simultaneous operation of the two ionization sources was investigated by recording ion mobility spectra of selected samples. The design allows one to observe peaks from either the corona discharge or photoionization individually or simultaneously. This makes it possible to accurately compare peaks in the ion mobility spectra from each individual source. Finally, the instrumentʹs capability for discriminating two peaks appearing in approximately identical drift times using each individual ionization source is demonstrated.
Keywords
ion mobility , Corona discharge , UV photoionization , Ionization source