• Title of article

    On-line measurement of perchlorate in atmospheric aerosol based on ion chromatograph coupled with particle collector and post-column concentrator

  • Author/Authors

    Takeuchi، نويسنده , , Masaki and Yoshioka، نويسنده , , Kaoru and Toyama، نويسنده , , Yusuke and Kagami، نويسنده , , Ai and Tanaka، نويسنده , , Hideji، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2012
  • Pages
    6
  • From page
    527
  • To page
    532
  • Abstract
    An automated analysis system has been developed for measuring perchlorate concentration in atmospheric aerosol. The perchlorate in aerosol sample, which has been collected with water mist in a hydrophobic filter/mist chamber based particle collector, is continuously preconcentrated. The matrix ions such as sulfate are subsequently removed from the preconcentrator. The remaining perchlorate is then analyzed on-line with an ion chromatograph in conjunction with a Nafion membrane tube based post-column concentrator. The sensitivity is increased by a factor of 7.7 with the post-column concentration technique. The proposed system has been successfully operated at Tokushima, Japan. The limit of detection is 0.35 ng/m3 for 3 h sampling cycle. The perchlorate concentration in the atmospheric aerosol averaged 1.01±1.75 ng/m3 (n=12).
  • Keywords
    atmospheric aerosol , automated analysis , Ion chromatography , Post-column concentration , Perchlorate
  • Journal title
    Talanta
  • Serial Year
    2012
  • Journal title
    Talanta
  • Record number