    Improvements in sensitivity and in anti-aging of Pt/C/Nafion® gases sensors modified by chromium

  • Author/Authors

    Liu، نويسنده , , Yu-Chuan and Hwang، نويسنده , , Bing-Joe and Tzeng، نويسنده , , Iou-Jeng، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    85
  • To page
    90
  • Abstract
    Pt/C/Nafion® composite electrodes prepared by a hot-pressed method were modified by chromium (Cr) for sensing hydrogen and oxygen in this study. The results indicate that the sensitivities of the Cr-modified electrodes in sensing hydrogen and oxygen rise markedly from 0.716 to 0.771 and from 0.414 to 0.533 μA per ppm, respectively, and the response and the recovery times on the Cr-modified electrode are shortened. Meanwhile, the Cr-modified Pt/C/Nafion® electrode can depress the aging in continuously sensing oxygen for 250 h. After aging, the sensitivities of the Pt/C/Nafion® electrode and the Cr-modified Pt/C/Nafion® electrode decrease by 81 and 65%, respectively. Generally, both the response and the recovery times decrease significantly with increasing gases concentrations. A sensing model was also proposed to illustrate the sensing phenomenon.
  • Keywords
    Chromium , Pt/C/Nafion® , Amperometric sensor
  • Journal title
    Journal of Electroanalytical Chemistry
  • Serial Year
    2002
  • Journal title
    Journal of Electroanalytical Chemistry
  • Record number