Title of article :
Application of response surface methodology (RSM) to the optimization of a post-column luminol chemiluminescence analysis of silyl peroxides
Author/Authors :
Baj، نويسنده , , Stefan and S?upska، نويسنده , , Roksana and Krawczyk، نويسنده , , Tomasz، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Abstract :
The possibility of the utilization of chemiluminescence post-column luminol oxidation (CL) in a HPLC system for silyl peroxides analysis has been investigated. The conditions of HPLC separation for 12 silyl peroxides, representing bissilyl and alkyl-silyl peroxides, as well as their potential impurities, were established. Optimal chemiluminescent post-column reaction conditions were found using central composite design (CCD) and response surface methodology (RSM). The interaction effects of four of the most important operating variables – the concentrations of luminol, hemin, sodium hydroxide and the post-column solution flow rate – on the light intensity were evaluated. The optimized conditions for analysis were the same for bissilyl and alkyl-silyl peroxides for the base concentration (0.03 M), the luminol concentration (0.4 g L−1) and the hemin concentration (0.3 g L−1). The only differences occurred in a reagent flow rate (for bissilyl peroxide −0.3 mL min−1 and for alkyl-silyl peroxides −0.9 mL min−1).
optimal conditions, the detection limits were in the 0.07–0.16 nM range for bissilyl, and 0.53–1.01 for alkyl-silyl peroxides. The calibration curves were linear in the 0.25–3 nM range for bissilyl and the 2.5–25 range for alkyl-silyl peroxides. Intra-day and inter-day precision was lower than 5.5% for each tested concentration level. A mechanism of luminol oxidation by silyl peroxides involving a hydrolysis step with the formation of hydrogen peroxide or hydroperoxide was proposed.
Keywords :
Chemiluminescence , Luminol , Determination , Silyl peroxides , Response surface methodology , HPLC