Title of article :
Relaxation and Simplex mathematical algorithms applied to the study of steady-state electrochemical responses of immobilized enzyme biosensors: Comparison with experiments
Author/Authors :
Flexer، نويسنده , , V. and Pratt، نويسنده , , K.F.E. and Garay، نويسنده , , F. Michael Bartlett، نويسنده , , P.N. and Calvo، نويسنده , , E.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
A description of the implementation of the relaxation method with automatic mesh point allocation for immobilized enzyme electrodes is presented. The advantages of this method for the solution of coupled reaction–diffusion problems are discussed. The relaxation numerical simulation technique is combined with the Simplex fitting algorithm to extract kinetic parameters from experimental data. The results of the simulations are compared to experimental data from self-assembled multilayered electrodes comprised of glucose oxidase (GOx) and an Os modified redox mediator and found to be in excellent agreement.
Keywords :
Relaxation Method , Simplex algorithm , Enzyme electrodes , Electrostatic self-assembly
Journal title :
Journal of Electroanalytical Chemistry
Journal title :
Journal of Electroanalytical Chemistry