Title of article :
Solid-contact Hg(II)-selective electrode based on a carbon-epoxy composite containing a new dithiophosphate-based ionophore
Author/Authors :
Juarez-Gomez، نويسنده , , J. and Perez-Garcia، نويسنده , , F. and Ramirez-Silva، نويسنده , , M.T. and Rojas-Hernandez، نويسنده , , A. and Galan-Vidal، نويسنده , , C.A. and Paez-Hernandez، نويسنده , , M.E.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Pages :
8
From page :
235
To page :
242
Abstract :
The experimental results herein concern a novel ion-selective electrode manufactured with an inner solid contact based on a new ligand O,O′-(2,2′-biphenylene)dithiophosphate pentyl (PenDTF). This electrode displays high selectivity toward the Hg(II) ion even in the presence of different divalent ions. The electrode also exhibited a good Nernstian response to Hg(II) (33.7±1.0 mV decade−1) over an ample concentration range (5.3×10−7−1.0×10−2 mol L−1), with a detection limit of (6.1±1.7)×10−7 mol L−1. As a result of design, the electrode can be used for many experiments, simply renewing its contact surface with emery paper, without giving rise to a significant response deviation, exhibiting a variation coefficient of 3%, capable of being tested within the 0–5 pH interval. The electrode was used satisfactorily as indicating electrode during the potentiometric titration of Hg(II) ions with EDTA.
Keywords :
Potentiometric titration , mercury (II) , ISE , Selectivity , ionophore