Title of article :
A high performance microfluidic analyser for phosphate measurements in marine waters using the vanadomolybdate method
Author/Authors :
Legiret، نويسنده , , François-Eric and Sieben، نويسنده , , Vincent J. and Woodward، نويسنده , , E. Malcolm S. and Abi Kaed Bey، نويسنده , , Samer K. and Mowlem، نويسنده , , Matthew C. and Connelly، نويسنده , , Douglas P. and Achterberg، نويسنده , , Eric P.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Pages :
6
From page :
382
To page :
387
Abstract :
We report a high performance autonomous analytical system based on the vanadomolybdate method for the determination of soluble reactive phosphorus in seawater. The system combines a microfluidic chip manufactured from tinted poly (methyl methacrylate) (PMMA), a custom made syringe pump, embedded control electronics and on-board calibration standards. This “lab-on-a-chip” analytical system was successfully deployed and cross-compared with reference analytical methods in coastal (south west England) and open ocean waters (tropical North Atlantic). The results of the miniaturized system compared well with a reference bench-operated phosphate auto-analyser and showed no significant differences in the analytical results (student’s t-test at 95% confidence level). The optical technology used, comprising of tinted PMMA and polished fluidic channels, has allowed an improvement of two orders of magnitude of the limit of detection (52 nM) compared to currently available portable systems based on this method. The system has a wide linear dynamic range 0.1–60 μM, and a good precision (13.6% at 0.4 μM, n=4). The analytical results were corrected for silicate interferences at 0.7 μM, and the measurement frequency was configurable with a sampling throughput of up to 20 samples per hour. This portable micro-analytical system has a low reagent requirement (340 μL per sample) and power consumption (756 J per sample), and has allowed accurate high resolution measurements of soluble reactive phosphorus in seawater.
Keywords :
nutrient analysis , Seawater , Microfluidic analyser , Silicate interference , Vanadomolybdate method , Phosphate measurement
Journal title :
Talanta
Serial Year :
2013
Journal title :
Talanta
Record number :
1668859
Link To Document :
بازگشت