Title of article
Large volume injection in gas chromatography using the through oven transfer adsorption desorption interface operating under vacuum
Author/Authors
Aragَn، نويسنده , , ءlvaro and Toledano، نويسنده , , Rosa M. and Gea، نويسنده , , Sara A. Cortes، نويسنده , , José M. and Vلzquez، نويسنده , , Ana M. and Villén، نويسنده , , Jesْs، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2014
Pages
6
From page
39
To page
44
Abstract
The present work describes a modification of the Through Oven Transfer Adsorption Desorption (TOTAD) interface, consisting of coupling a vacuum system to reduce the consumption of the helium needed to totally remove the eluent for large volume injection (LVI) in gas chromatography (GC).
fferent retention materials in the liner of the TOTAD interface were evaluated: Tenax TA, which was seen to be unsuitable for working under vacuum conditions, and polydimethylsiloxane (PDMS), which provided satisfactory repeatability as well as a good sensitivity. No variability was observed in the retention times in either case. Solutions containing organophosphorous pesticides in two different solvents, a polar (methanol/water) and a non-polar (hexane) solvent, were used to evaluate the modification.
cuum system coupled to the TOTAD interface allowed up to 90% helium to be saved without affecting the performance.
Keywords
Vacuum , Tenax , Through Oven Transfer Adsorption Desorption interface , polydimethylsiloxane , Large volume gas chromatography injection
Journal title
Talanta
Serial Year
2014
Journal title
Talanta
Record number
1670606
Link To Document