Title of article :
Trace metal determination in natural waters by automated solid phase extraction system and ICP-MS: The influence of low level Mg and Ca
Author/Authors :
Wang، نويسنده , , Bo-Shian and Lee، نويسنده , , Chih-Ping and Ho، نويسنده , , Tung-Yuan، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2014
Pages :
8
From page :
337
To page :
344
Abstract :
A fully automated high pressure pretreatment system with Nobias Chelate-PA1 resin (PA1) was developed for trace metal determination by ICP-MS in natural waters. By varying the concentrations of Mg and Ca to mimic the concentrations in the eluate obtained by PA1 or iminodiacetate type resins, the overall analytical performance of the system was assessed for the determination of Al, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Y, Cd, Ag, Pb and REE. Comparing with the low mM level Mg and Ca (both ranging from 1 to 4 mM) eluted by iminodiacetate type resins, the eluate obtained by PA1 contains sub-μM level Mg and Ca, which remarkably decrease matrix effect in ICP-MS analysis and significantly improve the analytical performance. With recovery better than 90% for most the trace metals examined, the accuracy was further verified through the analysis of five natural water reference materials with salinity spanning from 0 to 35‰. We have successfully applied the pretreatment system to determine trace metals in the seawater samples collected in the Western Philippine Sea through Taiwan GEOTRACES cruise.
Keywords :
Trace metal preconcentration , Iminodiacetate resins , Matrix effect , GEOTRACES , Nobias Chelate-PA1
Journal title :
Talanta
Serial Year :
2014
Journal title :
Talanta
Record number :
1671394
Link To Document :
بازگشت