• Title of article

    Selective and sensitive detection of nitrite based on NO sensing on a polymer-coated rotating disc electrode

  • Author/Authors

    Milczarek، نويسنده , , Grzegorz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    6
  • From page
    199
  • To page
    204
  • Abstract
    A new effective way of nitrite detection in complex samples is presented. It is based on chemical conversion of nitrite to nitric monoxide (NO) in acidic aqueous solution containing hexacyanoferrate(II) as a reductor. NO is then detected on a poly-eugenol coated platinum electrode. When the electrode is rotating and the reduction medium is continuously purged with nitrogen, the addition of a nitrite-containing sample produces narrowed current spikes. The peak current is proportional to nitrite content in the sample over the range of 5.0–100 μM and detection limit is 0.6 μM. The method is simple and highly reproducible. Relative standard deviation of 10 repetitions is less than 4%. Practical utility of the proposed approach is demonstrated by nitrite determination in human saliva.
  • Keywords
    Rotating disc electrode , nitrite , Nitric oxide , Poly-eugenol
  • Journal title
    Journal of Electroanalytical Chemistry
  • Serial Year
    2007
  • Journal title
    Journal of Electroanalytical Chemistry
  • Record number