Abstract :
Theory of chronoamperometry on ring-recessed microelectrodes and their arrays is presented. Analysis of the current transients measured at thin rings and their arrays shows that chronoamperometry at such devices is an excellent method of simultaneous measurement of the diffusion coefficient D and the product n[A], where n is the number of electrons transferred and [A] is the concentration of electroactive analyte. A generic, accurate and easy to use method of experimental chronoamperometric data analysis is proposed. It is shown that the method can be applied to the simultaneous measurement of D and n[A] in solution. The method is shown to be applicable to an analysis of data recorded at single disk and ring electrodes as well as arrays of these electrodes even in the case of diffusional non-independence of electrodes in the array.
Keywords :
Ring electrodes , Microring electrodes , Microelectrode array , microdisk , Measurements of diffusion coefficients
Journal title :
Journal of Electroanalytical Chemistry
Serial Year :
2010
Journal title :
Journal of Electroanalytical Chemistry
Record number :
1674520
Link To Document :
بازگشت