• Title of article

    Flow injection fluorimetric determination of chromium(VI) in electroplating baths by luminescence quenching of tris(2,2′-bipyridyl) ruthenium(II)

  • Author/Authors

    Hassan، نويسنده , , Saad S.M. and Abdel-Shafi، نويسنده , , Ayman A. and Mohammed، نويسنده , , Ayman H.K. Mohamed، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    696
  • To page
    702
  • Abstract
    A sensitive and selective luminescence quenching method is developed and used for manual and flow injection analysis (FIA) of chromium(VI) by reaction with [Ru(bpy)3]2+. The emission peak of ruthenium(II) at 595 nm is linearly decreased as a function of Cr(VI) concentration. This permits determination of chromium(VI) ion over the concentration range 0.1–20 μg ml−1 with a detection limit of 33 ng ml−1. The quenching process is due to an electron transfer from the luminescent [Ru(bpy)3]2+ complex ion to Cr(VI) resulting in the formation of the non-luminescent [Ru(bpy)3]3+ complex ion. Selectivity for Cr(VI) over many anions and transition, alkali and alkaline earth metal cations is demonstrated. High concentration levels of sulphate, chloride, borate, acetate, phosphate, nitrate, cyanide, Pb2+, Zn2+, Hg2+, Cu2+, Cd2+, Ni2+ and Mn2+ ions are tolerated. The effects of solution pH and [Ru(bpy)3]2+ reagent concentration are examined and the reaction conditions are optimized. Validation of the method according to the quality assurance standards show suitability of the proposed method for use in the quality control assessment of Cr(VI) in complex matrices without prior treatment. The method is successfully applied to determine chromium(VI) in electroplating baths using flow injection analysis. Results with a mean standard deviation of ±0.6% are obtained which compare fairly well with data obtained using atomic absorption spectrometry.
  • Keywords
    Flow injection analysis , Electroplating chromium waste , Ruthenium tris bipyridyl , Determination of chromium(VI) , Luminescence quenching
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number

    1674621