Title of article :
Sodium- and chloride-selective microelectrodes optimized for corrosion studies
Author/Authors :
Nazarov، نويسنده , , Valentine A. and Taryba، نويسنده , , Maryna G. and Zdrachek، نويسنده , , Elena A. and Andronchyk، نويسنده , , Kseniya A. and Egorov، نويسنده , , Vladimir V. and Lamaka، نويسنده , , Svetlana V.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
New sodium and chloride ion-selective membrane cocktails for glass-capillary microelectrodes were developed satisfying the demands for corrosion studies. Membrane cocktails were optimized to provide a wide pH working range, good long-term stability and reproducibility of potential. The proposed sodium-selective cocktail contains sodium ionophore VI, potassium tetrakis(4-chlorophenyl) borate and 2-nitrophenyl octyl ether. The chloride-selective cocktail consists of tridodecylmethylammonium chloride and 2-nitrophenyl octyl ether. The proposed microelectrodes were successfully applied to measure the concentration of Cl− and Na+ over corroding cut-edge samples of metallic coated steel and can be recommended for measuring the distribution of Na+ and Cl− in corrosion applications.
Keywords :
Localized electrochemical technique , SIET , Corrosion , Micro-potentiometry , Svet , Ion-selective microelectrode
Journal title :
Journal of Electroanalytical Chemistry
Journal title :
Journal of Electroanalytical Chemistry