Title of article :
Optimization of hollow cathode diameter for particle beam/hollow cathode glow discharge atomic emission spectrometry
Author/Authors :
Dempster، نويسنده , , Melissa A. and Marcus، نويسنده , , R.Kenneth، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
12
From page :
599
To page :
610
Abstract :
An evaluation of the effect of cathode diameter on analyte emission intensity for the particle beam/hollow cathode glow discharge atomic emission spectroscopy (PB–HC–AES) system is described. A high-efficiency thermoconcentric nebulizer is used to generate a fine aerosol of the aqueous solution flowing at rates of ∼1 ml/min. Desolvation and use of a momentum separator/particle beam interface serves to introduce analyte particles into a heated HC glow discharge source for subsequent vaporization, atomization and excitation. The effects of discharge gas (He) pressure and current on emission responses for sodium and copper are evaluated in flow injection mode using 200-μl injection volumes for a series of fixed-length (25 mm) cylindrical HC having inner diameters of 2, 3, 5 and 7 mm. Analytical response curves obtained with the 2-mm i.d. HC for sodium and copper yield detection limits of 0.6 and 1.3 ppb in elemental Na and Cu, respectively, with less than 10% relative standard deviation (R.S.D.) for triplicate injections over a concentration range of 0.5–20 ppm.
Keywords :
particle beam , Hollow cathode diameter , Atomic emission , Glow discharge , Liquid chromatography
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2000
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1678525
Link To Document :
بازگشت