Title of article :
Calibration in flame atomic absorption spectrometry using time-dependent concentration profiles
Author/Authors :
L?pez-Garc??a، نويسنده , , Ignacio and Vi?as، نويسنده , , Pilar and Gonz?lvez، نويسنده , , José and Hern?ndez-C?rdoba، نويسنده , , Manuel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A computer-controlled manifold allowing the automatic collection of concentration profiles for use in flame atomic absorption spectrometry (FAAS) is described. For this purpose a standard solution is pumped at an increasing flow rate, the difference between this flow rate and the nebulizer uptake rate being compensated with water. Next, the sample solution is pumped at a decreasing rate. A 50-s gradient time is used. The absorbance–time profiles thus obtained intersect at a point which permits the concentration to be calculated. The system allows the on-line dilution of solutions which are too concentrated to be measured by direct aspiration. Using a 30 μg ml−1 calcium standard solution for calibration, this analyte can be determined in the 1–200-μg ml−1 range with relative errors below ±1%.
Keywords :
Atomic absorption spectrometry , Automation , Calibration , On-line dilution
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy