Title of article :
Validation of an X-ray methodology with environmental concern
Author/Authors :
Morgenstern، نويسنده , , P. and Brüggemann، نويسنده , , L. and Wennrich، نويسنده , , R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
The capabilities of test procedures toward the validation of X-ray fluorescence analysis with environmental concern were investigated. The applied analytical scheme was based on the combination of fusion and pelletizing technique for sample preparation, while matrix correction was performed by means of the fundamental parameter and the scattered radiation method. Validation was carried out in terms of precision, trueness, measurement uncertainty, limit of detection and test for homogeneity of the sample material. To confirm trueness for the examined calibration strategy, the bias of the recovery functions for each analyte under study was evaluated by means of orthogonal regression analysis. The investigations were focused on single laboratory validation. Consequently, the presented validation procedures were based nearly exclusively on certified reference material.
Keywords :
Method validation , Precision studies , Estimation of trueness , Environmental samples , X-ray fluorescence analysis
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy