• Title of article

    Validation of an X-ray methodology with environmental concern

  • Author/Authors

    Morgenstern، نويسنده , , P. and Brüggemann، نويسنده , , L. and Wennrich، نويسنده , , R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    13
  • From page
    185
  • To page
    197
  • Abstract
    The capabilities of test procedures toward the validation of X-ray fluorescence analysis with environmental concern were investigated. The applied analytical scheme was based on the combination of fusion and pelletizing technique for sample preparation, while matrix correction was performed by means of the fundamental parameter and the scattered radiation method. Validation was carried out in terms of precision, trueness, measurement uncertainty, limit of detection and test for homogeneity of the sample material. To confirm trueness for the examined calibration strategy, the bias of the recovery functions for each analyte under study was evaluated by means of orthogonal regression analysis. The investigations were focused on single laboratory validation. Consequently, the presented validation procedures were based nearly exclusively on certified reference material.
  • Keywords
    Method validation , Precision studies , Estimation of trueness , Environmental samples , X-ray fluorescence analysis
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2004
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number