Title of article :
High intensity monocapillary X-ray guide tube with 10 micrometer spatial resolution for analytical X-ray microscope
Author/Authors :
Ohzawa، نويسنده , , Sumito and Komatani، نويسنده , , Shintaro and Obori، نويسنده , , Kenich، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
High-intensity monocapillary X-ray guide tube is studied and compared with a commercially available polycapillary tube. The newly designed monocapillary tube is capable of achieving spatial resolution of 10 μm in beam diameter. In comparison with the polycapillary tube, beam diameter of the present monocapillary tube is not enlarged and allows the sample to be measured at a working distance that is 10 times less. Thus, the monocapillary tube is the more versatile X-ray focusing optics as it provides improvements in spatial analysis of samples.
Keywords :
X-ray guide tube , Analytical X-ray microscope , Monocapillary , Polycapillary
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy