Title of article :
A hydride generation flow system for determination of arsenic and selenium by total reflection X-ray fluorescence spectrometry
Author/Authors :
Gomes، نويسنده , , Ana Carla F. and Meneg?rio، نويسنده , , Amauri A. and Pellegrinotti، نويسنده , , Daniel C. and Giné، نويسنده , , Maria F. and Nascimento Filho، نويسنده , , Virg??lio F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
In this work, a preconcentration and separation system based on continuous flow hydride generation is proposed to improve the determination of As and Se by total reflection X-ray fluorescence spectrometry. The generated hydrides are continuously separated from the liquid phase and collected in a chamber containing 250 μl of HCl/HNO3 1:1 (v/v) solution. Hydride generation conditions and collection of the hydrides were evaluated. Under optimised conditions, enrichment factors of 55 for As and 82 for Se were attained. Detection limits of 0.3 μg l−1 for As and Se were obtained when 20 ml of sample was used. Analysis of a natural water standard reference material from National Institute of Standard and Technology (SRM-1640) was in agreement with the certified values at the 95% confidence level.
Keywords :
Hydride generation , TXRF , Preconcentration , Flow system
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy